Neonatal Skin Temperature Probes Premature Skin Temperature Probes Baby incubators - Copyright 2005 - All rights reserved EXACON SCIENTIFIC

 
DK - DE - FR - IT - ES - RU
Neonatal Skin Temperature Probes
High Quality Reusable Neonatal Skin Temperature Probes and Sensors for
Neonatal and Premature Temperature Monitoring in OP - ICU - Pediatrics - Baby Incubators
 
The D-S06 Series Neonatal Skin Temperature Probes are extremely small and designed for  measurement on
thorax, axilla and extremities of neonates and prematures in OP - ICU - Pediatrics - Baby Incubators
 
The probes are available for all brands of modern patient monitors and baby incubators
 
Product description

Connectors
As standard available with the type A, JK, HP and SA. - Other connector types are available upon request.  
 
Cables
3 mm black medical grade silicone tubing, with free inner shielded cable. Primary standard length 2,75 m - Secondary standards 1,5 m and 4,5 m. Other lengths are available upon request.
 
Electrode
ø2 x450 mm electrode cable with solid stainless steel disk, ø6 x 2 mm.  The sensor chip is placed in the center of the disk giving an optimized thermal surface contact and a minimum influence from ambience.  
 
Measuring reference
2,252 kohms at 25ºC.
 
Measuring accuracy
  
0...25ºC ±0,2ºC
25...50ºC ±0,1ºC
 
Self heating
<
0,01ºC/45µW
 
Insulation
Biological saline solution minimum. 2,1kVp 50Hz
 
Use and Maintenance
The temperature probe can be cleaned with ordinary detergents either by wiping it clean with a soft cloth or machine washing
at e.g.  80ºC ~ 176ºF.
 
Disinfecting
Without changing the features of the soft and smooth structure, the probes are accepting all generally applied  hospital disinfectants
such as ethanol, isopropyl alcohol or chlorite compounds.
 
Typical examples
Disinfect the cleaned probe with a disinfectant such as  
A. Rodalon™, Cidex™.
B. 1 - 2,5% Formaldehyde solution.
C. 70% Alcohol, 0,5% Chlorexidine, 29,5% H2O
 
 
Steam sterilization
Upon request
 
Standards
EN12470-4, EN60601-1, EN10993
 
CE-marking
Approved in Class IIb to Council Directive 93/42/EEC Annex II,  section 3.2
Full quality assurance system EN ISO 9001 and EN ISO13485
 EXACON™