Tympanic Membrane Temperature Probes
 
Central body temperature registration by measurement of the ear temperature
 
The D-TM1 probe series Tympanic Membrane Temperature Probes and Sensors are designed for measurement of the inner ear temperature, by physical sensor contact to the Tympanic Membrane.
This method to measure the ear temperature produces a fast and close correlation to the body core temperature, and it is in  particular used during anesthesia to identify malignant hyperthermia at a earliest possible stage. Also used by Emergency Teams at accidental hypothermia.
 
Changes of the body temperature is recorded immediately by measurement of the ear temperature. All other places for central temperature measurement, such as esophagus, nasopharynx and rectum are reacting slowly with changes of the  body temperature regulation at the pituitary gland.
 
The tympanic membrane probe is inserted in the cleaned ear and kept in firm position by a piece of cotton around the electrode wire.
Outside the ear, the wire is taped to the skin to assure a stable positioning of the temperature sensor tip. Insertion must be performed with care, as the tympanic membrane is thin and woundable. The strength of the tympanic membrane can be very variable.
 
The measuring point of the tympanic membrane temperature probes is shaped as a small ball with a diameter of approximately 1 mm. It is located in the tip of a 30 cm long electrode wire made by a very soft and flexible silicone hose with a diameter about 1 mm.
 
The electrode wire of the tympanic membrane probe is in a firm assembly junction connected to a silicone cable with the diameter
3,0 mm and a length of 2,75 m.
The tympanic membrane probes and sensors are available with the most important connector types “JK”, “A”, “HP”, "SA".
Other connectors are available upon request.
 
 
Measuring reference
2,252 kohms at 25ºC.
 
Measuring accuracy
  
0...25ºC ±0,2ºC
25...50ºC ±0,1ºC
 
Self heating
<
0,01ºC/45µW
 
Insulation
Biological saline solution minimum. 2,1kVp 50Hz.
 
Use and Maintenance

The temperature probe can be cleaned with ordinary detergents either by wiping it clean with a soft cloth or machine washing
at e.g.  80ºC ~ 176ºF.
 
Disinfecting
Without changing the features of the soft and smooth structure, the probes are accepting all generally applied  hospital disinfectants
such as ethanol, isopropyl alcohol or chlorite compounds.
 
Typical examples
Disinfect the cleaned probe with a disinfectant such as  
A. Rodalon™, Cidex™.
B. 1 - 2,5% Formaldehyde solution.
C. 70% Alcohol, 0,5% Chlorexidine, 29,5% H2O
 
 
Steam sterilization
A, SA and HP-types upon request  
Standards
EN12470-4, EN60601-1.

CE-marking
Approved in Class IIb to Council Directive 93/42/EEC Annex II,  section 3.2
Full quality assurance system EN ISO 9001 and EN ISO13485