Oral Nasal Temperature Probes, Oral Nasal Temperature Sensors, Oral Temperature, Nasal Temperature, Copyright 2005 All rights reserved EXACON SCIENTFIC

 
DK - DE - FR - IT - ES - RU
Oral & Nasal Temperature Probes
High Quality Reusable Oral Temperature Probes & Nasal Temperature Probes and Sensors
 With Ultra Fast Response Time - Flexible Teflon Electrode With Small Diameter
 
The D-F1310 Series Oral and Nasal Temperature Probes and Sensors are designed for a large variety
of applications, like fast oral  temperature registration in emergency room and for long term
nasal temperature measurements in OP, ICU and Pediatrics.
 
The  Oral and Nasal Temperature Probes and Sensors are available for all brands of modern patient monitors
 
Product description

Connectors
As standard available with the type A, JK, HP and SA. - Other connector types are available upon request.  
 Cables
3
mm black medical grade silicone tubing, with free inner shielded cable.
The primary standard length 2,75 m - Secondary standards 1,5 m and 4,5 m. Other lengths are available upon request.
 
Electrode
The white colored flexible plastic catheter has a nom. ø1,2 mm and length 100mm.
The sensor chip is placed in the very tip with a nominal outside diameter of 1,3 mm.
 
Measuring reference
2,252 kohms at 25ºC.
 
Measuring accuracy
  
0...25ºC ±0,2ºC
25...50ºC ±0,1ºC
 
Self heating
<
0,01ºC/60µW
 
Insulation
Biological saline solution minimum. 2,1kVp 50Hz.
 
Use and Maintenance
The temperature probe can be cleaned with ordinary detergents either by wiping it clean with a soft cloth or machine washing
at e.g. 80ºC ~ 176ºF.
 
Disinfecting
Without changing the features of the soft and smooth structure, the probes are accepting all generally applied  hospital disinfectants
such as ethanol, isopropyl alcohol or chlorite compounds.
 
Typical examples
Disinfect the cleaned probe with a disinfectant such as  
A. Rodalon™, Cidex™.
B. 1 - 2,5% Formaldehyde solution.
C. 70% Alcohol, 0,5% Chlorexidine, 29,5% H2O  
 
Steam sterilization
A, SA and HP-types upon request
 
Standards
EN12470-4, EN60601-1.
 
CE-marking
Approved in Class IIb to Council Directive 93/42/EEC Annex II,  section 3.2
Full quality assurance system EN ISO 9001 and EN ISO13485
   EXACON™