Standard and Autoclavable Temperature Probes
and Autoclavable Temperature Sensors
for Medical Use
Use
and Maintenance
-
EXACON Autoclavable Temperature probes are
designed for Steam Autoclave
Sterilization at temperatures up to 134°C.
- Using Central Temperature Probes
Introduce probe,
e.g. into the esophagus, the nasopharynx, the auditory canal
up to the tympanic membrane, or the rectum. Use only water
based crèmes for introduction.
-
- Nasopharyngeal application
The correct
position has been reached if you feel resistance when trying
to advance the probe further.
- Caution! Introduction of the
probe may induce nose bleeding
-
- Esophageal application
Introduce the
probe until it is situated in the inferior quarter of the esophagus.
-
- Aural application
Advance the probe until tip reaches
the tympanic membrane. Wrap a layer of cotton wool around the
cable to close the auditory canal.
- Wind the electrode wires with
a couple of turns around the outer ear to reduce heat transmission
to ambient.
-
Rectal application The probe is in correct position when it has
passed the sphincter. Use the disposable protective sheet
type PC1 - It is LATEX free and have a thin layer of Vaseline
for easiest introduction.-
PC1 
-
- Skin and Extremity Temperature
Probes
Attach probe
to finger or toe, using adhesive tape. Take care not to compress
blood vessels. Furthermore secure lead
- with adhesive tape. - Readout
stabilizes within approximately 3 min.
- Notice
Different parts of the body may
differ significantly in temperature.
- Monitoring temperature from
two different sites, e.g. oesophagus and extremity, provides
indirect information on the blood flow.
-
-
Cleaning Temperature Probes
Remove adhesive tape, use cleaning benzine to remove rests of
adhesive. Clean with a cloth dampened with soap water.
Caution! Do not immerse connector in the cleaning solution.
The temperature probe can be cleaned with ordinary detergents
either by wiping it clean with a soft cloth or machine washing
at e.g. 80ºC
~ 176ºF.
Disinfecting
Temperature Probes Without
changing the features of the soft and smooth structure, the
probes are accepting all generally applied hospital disinfectants
such as ethanol,
isopropyl alcohol or chlorite compounds. Typical examples Disinfect the cleaned probe with a disinfectant
such as A. Rodalon™, Cidex™. B. 1 - 2,5% Formaldehyde
solution. C. 70% Alcohol, 0,5% Chlorexidine, 29,5% H2O
Steam Sterilization
of Temperature Probes :
EXACON autoclavable
temperature probes accept steam temperature up to 134ºC according the the attached User's Instruction
Chemical Sterilization
With an ethylene
oxide mixture.
Other sterilization
systems
-
-
|
EXACON™
Placement
of Medical Temperature probes

Cleaning

Disinfecting

Autoclaving

|